Clinical comparison of three SARS-CoV-2 nucleic acid amplification tests for routine diagnostic testing (2024)

  • Journal List
  • Heliyon
  • v.9(11); 2023 Nov
  • PMC10685266

As a library, NLM provides access to scientific literature. Inclusion in an NLM database does not imply endorsem*nt of, or agreement with, the contents by NLM or the National Institutes of Health.
Learn more: PMC Disclaimer | PMC Copyright Notice

Clinical comparison of three SARS-CoV-2 nucleic acid amplification tests for routine diagnostic testing (1)

Link to Publisher's site

Heliyon. 2023 Nov; 9(11): e22112.

Published online 2023 Nov 7. doi:10.1016/j.heliyon.2023.e22112

PMCID: PMC10685266

PMID: 38034696

Tetiana Garmatiuk,a, Christine Gränitz-Trisko,a Charlotte Sochor-Geischläger,a Theresa Polsterer,a Francesca Caselotto,a Lukas Willitsch,a Birgit Reinhardt,b and Wolfgang Hufc

Author information Article notes Copyright and License information PMC Disclaimer

Associated Data

Data Availability Statement

Abstract

Background

Cycle threshold (Ct) values from SARS-CoV-2 nucleic acid amplification tests have been used to estimate viral load for treatment decisions. Additionally, there is a need for high-throughput testing, consolidating a variety of assays on one random-access analyzer.

Objectives

In this study, the clinical performance of the Alinity m SARS-CoV-2, RealTime SARS-CoV-2, and GeneXpert Xpress SARS-CoV-2/Flu/RSV assays was assessed.

Methods

Alinity precision and detection rates were evaluated using a dilution series of the Alinity m SARS-CoV-2 positive control. In a retrospective study, 7 remnant external quality assessment (EQA) specimens and 200 remnant nasopharyngeal swab specimens (100 positive and 100 negative) were tested in the three assays.

Results

Alinity had 100% detection rate at 50 copies/mL and high reproducibility (Ct value coefficient of variation ≤3.1%). All three assays correctly detected positive and negative EQA samples with comparable Ct values (max difference 2.38) and high linearity. In patient samples, positive percent agreement was 95% (95% CI 89–98%) and negative percent agreement was 100% (95% CI 96–100%) for Alinity, compared to the other two assays. Four specimens detected on Alinity m but not RealTime or Xpert had Ct values above 40. Assay results were highly correlated (r≥0.94). Ct values (after addition of 10 unread cycles to the reported Ct of RealTime) were comparable across the three assays.

Conclusions

Alinity m had high precision and accuracy and Ct values comparable to those of the RealTime and Xpert assays. The assays could be used interchangeably, with no need for adjustment of patient management decisions based on Ct values from each assay.

Keywords: SARS-CoV-2, Real-time PCR, Alinity m, Respiratory viruses

Highlights

1. Introduction

During the COVID-19 pandemic, the cycle threshold (Ct) value of SARS-CoV-2 nucleic acid amplification tests (NAATs) gained prominence as a way to semi-quantify viral load and guide patient management, despite continuing controversy regarding its utility [[1], [2], [3], [4], [5]]. A Ct value of 30 has been used to make clinical decisions on the duration of patient quarantine or discharge from hospital in absence of clinical symptoms [6]; however, intra-assay variability between multiple target gene regions of the test methods, combined with inter-assay variability, present challenges for the clinical interpretation of Ct values and their application to decision making [[7], [8], [9]].

We evaluated the Ct values and clinical performance of the recently launched Abbott Alinity m SARS-CoV-2 assay compared to two SARS-CoV-2 RT-PCR assays currently used in our laboratory for routine testing of all patients admitted to the hospital (Abbott RealTime SARS-CoV-2) or for urgent sample testing (Cepheid GeneXpert Xpress SARS-CoV-2/Flu/RSV).

2. Methods

2.1. Study design

This retrospective study was performed in the Department of Laboratory Medicine, Hietzing Hospital, Vienna, Austria. All clinical samples were deidentified prior to use in the study.

2.2. NAAT assays

The Alinity m SARS-CoV-2 assay (Abbott Molecular Inc., Des Plaines, IL, USA; “Alinity”) is indicated for the qualitative detection of SARS-CoV-2 in patients with or without signs and symptoms of COVID-19 infection. The assay can be run with nasopharyngeal, oropharyngeal, or nasal swabs, including pooled samples and saliva specimens [10]. Alinity is run on the Alinity m system, a fully automated, random-access analyzer with a time to first result of less than 2h that reports approximately 1000 results in 24h. The RealTime SARS-CoV-2 assay (Abbott Molecular Inc.; “RealTime”) is run on the m2000 batch analyzer with a turnaround time of approximately 6h and is indicated for the qualitative detection of SARS-CoV-2 in nasopharyngeal and oropharyngeal swabs from patients suspected of COVID-19 infection [11]. The GeneXpert Xpress SARS-CoV-2/Flu/RSV assay (Cepheid, Sunnyvale CA, USA; “Xpert”), run on the GeneXpert system with a low throughput and a turnaround time of approximately 1h, and is indicated for the qualitative detection and differentiation of SARS-CoV-2, influenza A, influenza B, and respiratory syncytial virus (RSV) viral RNA in nasopharyngeal swabs, nasal swabs, or nasal wash/aspirate specimens from patients suspected of respiratory viral infection [12]. All assays were performed according to the manufacturers’ instructions.

2.3. Analytical performance assessment

To assess Alinity precision and detection rates around the limit of detection (LOD) at 100 copies/mL [10], a dilution panel was prepared using the Alinity m SARS-CoV-2 positive control diluted in 0.9% NaCl solution to concentrations of 400, 300, 200, 100, and 50 copies/mL. Each panel member was tested in 20 replicates; the 100 copies/mL dilution was tested in 40 replicates. Assay precision was assessed by calculating means, standard deviations (SD), and coefficients of variation (CV) of Ct values. Detection rates were also calculated.

Seven leftover external quality assessment (EQA) specimens from the Austrian Association for Quality Assurance and Standardization of Medical and Diagnostic Tests (OEQUASTA), the national EQA provider in Austria, were tested retrospectively. The EQA panel comprised two diluted SARS-CoV-2 positive swab specimens (including one alpha variant), one SARS-CoV-2 positive viral culture supernatant sample, one negative control (0.9% NaCl solution), and one negative diluted swab specimen containing human RNA. The viral culture supernatant sample was tested undiluted and diluted (1:10 and 1:100) [13]. The leftover specimens had been stored at −80°C for 6 months prior to testing on the three assays.

2.4. Clinical comparison of the three assays

A total of 200 remnant nasopharyngeal swab specimens collected over the previous 4 months and stored at −25°C, with valid Ct values from one of the three assays, were analyzed with the other two assays. The samples were collected during monitoring of hospitalized patients and screening of newly admitted patients. The patient cohort included patients with respiratory symptoms as well as patients without respiratory symptoms that were admitted to the hospital for other reasons. For the analysis, 100 negative specimens and 100 positive specimens (Ct values 16.90–40.99) were randomly selected based on the Alinity results. As all remnant clinical specimens were anonymized prior to the study and no patient-specific information was collected, approval by an ethics committee was deemed unnecessary.

2.5. Data and statistical analysis

As Ct values are non-standardized and none of the assays could be considered the reference method, we used Pearson correlation to compare the analytical and clinical performance of the three assays. Statistical evaluation was conducted by Microsoft Excel and the add-in software Analyse-it.

3. Results

3.1. Analytical performance of alinity

The observed detection rates for all concentrations of the dilution panel were ≥97.5% (Table 1). Additionally, Alinity had high reproducibility, with a Ct value CV of ≤3.1%. In testing leftover EQA specimens, all positive and negative samples were correctly identified by all three assays (Table 2). Comparable Ct values were observed across the three assays with a maximum difference of 2.38. Linearity was high for the three dilution series samples, with Pearson's correlation coefficients ranging between −0.999 and −1.000 (data not shown).

Table 1

Detection Rates with the Alinity m SARS-CoV-2 Assay.

Target concentration (copies/mL)Replicates tested (n)Replicates detected (n)Percentage of replicates detectedMean Ct±SDCt CV%
4002020100%34.20±1.063.1%
3002020100%34.12±0.752.2%
2002020100%34.91±0.892.6%
100403997.5%36.13±0.872.4%
502020100%37.78±1.002.7%

Open in a separate window

Ct: cycle threshold; CV: coefficient of variation; SD: standard deviation.

Table 2

Evaluation of leftover samples of the austrian association for quality assurance and standardization of medical and diagnostic tests (OEQUASTA) [13].

EQA SpecimenDilution SeriesViral Load (copies/mL)Expected ResultAlinity Result
(Ct value)
RealTimea
Result
(Ct value)
Xpert Result
(Ct value)
Human RNAnegnegnegneg
SARS-CoV-2∼252,000pos25.5226.5127.90
SARS-CoV-21:10∼22,300pos31.3431.7032.50
SARS-CoV-2
(alpha)
∼345,300pos26.7628.3528.90
0.9% NaClnegnegnegneg
SARS-CoV-21:1∼96,100pos28.9529.4130.50
SARS-CoV-21:100∼640pos36.0036.7338.30

Open in a separate window

EQA: external quality assessment; pos: positive result; neg: negative result.

a10 unread cycles included.

3.2. Clinical performance of alinity against comparators

In testing 200 clinical samples, Alinity had a positive percent agreement (PPA) of 95% (95% CI 89–98%) and a negative percent agreement (NPA) of 100% (95% CI 96–100%; Table 3) compared to RealTime and Xpert. Of the 100 positive specimens, 94 yielded positive results with all three assays. Of the remaining 6 specimens, 4 samples with Alinity Ct values above 40 were negative on RealTime and Xpert. Two Alinity positive specimens with Ct values of 39.04 and 40.76 were found to be positive with either one of the other comparator assays. All 100 negative samples with Alinity were also negative with RealTime and Xpert.

Table 3

Agreement between alinity, RealTime, and xpert assays.

Alinity +
RealTime +
Xpert +
Alinity +
RealTime +
Xpert -
Alinity +
RealTime -
Xpert +
Alinity +
RealTime -
Xpert -
Alinity -
RealTime -
Xpert -
Alinity/RealTime AgreementAlinity/Xpert AgreementRealTime/Xpert Agreement
Alinity+(n=100)941a1b4cPPA95%
(CI 89–98%)
95%
(CI 89–98%)
99%
(CI 94–100%)
Alinity - (n=100)100NPA100%
(CI 96–100%)
100%
(CI 96–100%)
99%
(CI 95–100%)

Open in a separate window

PPA: Positive percent agreement; NPA: Negative percent agreement; CI: 95% confidence interval.

aCt (Alinity): 39.04.

bCt (Alinity): 40.76.

cCt (Alinity): 40.41–40.99.

The three assays showed high correlation (p<0.0001), with r values of 0.95 between Alinity and RealTime, 0.94 between Alinity and Xpert, and 0.96 between RealTime and Xpert (Fig. 1A–C). With addition of the 10 unread cycles to the reported RealTime Ct values, Bland-Altman evaluations showed low mean Ct differences between the assays: 2.55 (Alinity – RealTime), 1.46 (Alinity – Xpert), and 1.10 (Xpert – RealTime). Individual specimen Ct values had a mean difference of 1.7±0.1 (SE).

Open in a separate window

Fig. 1

Correlation of Ct values for (A) Alinity and RealTime, (B) Alinity and Xpert, and (C) Xpert and RealTime. Ct values for RealTime include the 10 unread cycles.

4. Discussion

We found that Alinity had a high precision with a Ct value CV of ≤3.1%. This observation was similar to previous studies that reported Ct value CVs of ≤1.9% [14,15]. In our study, the detection rate of 100% at 50 copies/mL by Alinity exceeded the sensitivity claimed by the manufacturer. Our finding was concordant with other assessments of the LOD of Alinity using similar dilution panels and observing LODs of 50–100 copies/mL [[14], [15], [16]].

When testing leftover Austrian OEQUASTA EQA specimens including the SARS-CoV-2 variant alpha (B.1.1.7), we observed high accuracy of all three assays, matching the expected results. Ct values (after addition of the 10 unread cycles to the reported Ct of RealTime) were comparable across the three assays, with a maximum Ct difference of 2.38 across assays and samples, and showed high linearity. Of note, across all participants in the EQA with a variety of platforms, the reported differences between the highest and the lowest Ct values within the same samples ranged between 12.7 and 17.8, and some laboratories did not report linearity in the dilution series [13]. Our results confirmed the accurate detection of multiple SARS-CoV-2 variant lineages, including B.1.1.7, B.1.351, and P.1 variants, by Alinity and RealTime as demonstrated previously by testing dilution series of heat-inactivated virus cultures down to concentration levels of 0.73–2.85 log copies/mL. In that study, all dilutions across the different variants were reliably detected by both assays [17].

In our performance assessment, 100 samples were tested negative with all three assays resulting in 100% negative percent agreement and suggesting high specificity for the assays evaluated in the study. This is important to correctly identify individuals without SARS-CoV-2 infection [18].

Significant and high correlation was seen across all three assays in our study, with correlation coefficients ≥0.94. Notably, the three assays produced similar Ct values, with mean Ct differences ranging between 1.10 and 2.55. This is of specific interest because in a recent study, the range of obtained Ct values across all evaluated test methods was approximately 30 Ct values per sample, respectively. In that study, two reference samples of about 6 and 7 Log copies/mL had been tested in 305 laboratories using overall 1109 test methods. Thus, for optimal interpretation of positive test results, it was recommended to anchor Ct values of individual test systems to reference materials of known viral RNA concentrations [9]. Similar to our results, a high correlation between Ct values of Alinity and RealTime (R2=0.95) and a mean difference of 4.14 Ct values were previously observed if including the unread cycles for RealTime [16]. Another study compared Alinity with other fully automated SARS-CoV-2 assays, Roche cobas® SARS-CoV-2 (cobas® 6800) and GeneXpert Xpress SARS-CoV-2, as well as with two semi-automated SARS-CoV-2 assays by Primerdesign (genesig®) and R-Biopharm (RIDA®GENE), respectively. The Ct values of the four assays correlated well with Alinity (r≥0.946) while the CVs of the Ct values ranged from 0.7% to 3.2% for the fully automated assays compared to 5.8–6.6% for the semi-automated tests, suggesting a higher precision of the fully automated assays. The median Ct values had a difference from Alinity of >7.6 Ct for the semi-automated assays compared to a difference of 3.1 Ct for cobas and 5.1 Ct for GeneXpert [19].

The results of our study suggest a slightly higher sensitivity of Alinity compared to RealTime and Xpert. Similarly, a comparison of Alinity, RealTime, and Seegene Allplex SARS-CoV-2 showed 100% NPA and 100% PPA while one of the two targets of an inhouse assay appeared to be slightly less sensitive [14]. A higher sensitivity was also observed for Alinity compared to Roche cobas SARS-CoV-2 (cobas 6800) as 37 out of 998 samples positive by Alinity were not detected by cobas 6800 while all 961 samples positive by cobas 6800 were also positive with Alinity [20]. Moreover, when testing a dilution series of a SARS-CoV-2 positive cell culture supernatant, Alinity was one of the two assays with the highest sensitivity among multiple commercial SARS-CoV-2 assays tested [21]. Finally, high concordance was also observed when comparing Xpert to Alinity m Resp-4-Plex (detecting SARS-CoV-2, influenza A/B, RSV) and Roche cobas Liat SARS-CoV-2/influenza A/B. Alinity m Resp-4-Plex exhibited the lowest LOD for SARS-CoV-2 at 26 copies/mL compared to 58 copies/mL for cobas Liat and 83 copies/mL for Xpert [22]. As many laboratories use multiple test methods for SARS-CoV-2 detection, it is important to understand how individual assays compare to each other.

Some authors suggested that results with high Ct values are potentially false positive results. This assumption is based on the observation that follow-up samples collected within 72h or longer could confirm initial Alinity m results >37 Ct only in 13/40 (32.5%) of cases and initial in-house assay results with 34–40 Ct only in 11/339 (3.2%) of cases. However, as the median time interval between the index test and a positive follow-up test result was less than 1 day (22h), the positivity rate of the follow-up sample may have been impacted by the interval length between the collection dates of the two samples [23]. Clinical significance of SARS-CoV-2 results with high Ct values should be interpreted in the context of the clinical history of a patient. Detection of viral RNA does not necessarily represent transmissible virus, and the relationship between timing of SARS-CoV-2 detection and duration of infectivity is not fully understood. SARS-CoV-2 detection is considered critically important in the early phase of infection when patients are highly infectious and need to be quarantined. In contrast, prolonged virus detection beyond day 10 after symptom onset or day of positivity [24] is likely to be of minor clinical relevance with a low risk of transmission expected [25].

Limitations of this study were that the detection rates were evaluated before the 1st WHO International Standard for SARS-CoV-2 RNA was available. Thus, the use of the Alinity m SARS-CoV-2 positive control to prepare the dilution panel may limit the ability to compare our findings to other SARS-CoV-2 assays. Furthermore, additional panel members with lower concentrations would have provided a better estimation of the actual LOD in this study. However, in previous studies, detection rates for viral RNA less than 50 copies/mL dropped below 95% [14,15]. Samples with positive results on Alinity but negative on one or both of the two comparator assays could not be re-run due to limited sample volume.

5. Conclusions

Our study demonstrated that results from Alinity, run on the random-access Alinity m analyzer, correlated well with results from RealTime and Xpert in detecting SARS-CoV-2, providing similar Ct values. Thus, it can be used interchangeably as part of the clinical routine.

Credit author statement

Tetiana Garmatiuk; Wolfgang Huf, Dr: Conceived and designed the experiments; Analyzed and interpreted the data; Wrote the paper. Christine Gränitz-Trisko, Dr.: Contributed reagents, materials, analysis tools or data. Charlotte Sochor-Geischläger; Theresa Polsterer; Lukas Willitsch: Performed the experiments.Francesca Caselotto: Conceived and designed the experiments. Birgit Reinhardt, Dr.: Analyzed and interpreted the data; Wrote the paper.

Data availability statement

Data will be made available on request.

Additional information

No additional information is available for this paper.

Declaration of competing interest

The authors declare the following financial interests/personal relationships which may be considered as potential competing interests:Tetiana Garmatiuk reports article publishing charges and writing assistance were provided by Abbott Laboratories.

Acknowledgements

The authors would like to thank Stacey Tobin for her support in manuscript preparation.

Alinity m SARS-CoV-2 and RealTime SARS-CoV-2 testing kits used for this study were provided by Abbott Molecular Inc.

References

1. Rao S.N., Manissero D., Steele V.R., Pareja J. A systematic review of the clinical utility of cycle threshold values in the context of COVID-19. Infect. Dis. Ther. 2020;9:573–586. doi:10.1007/s40121-020-00324-3. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

2. Rao S.N., Manissero D., Steele V.R., Pareja J. Correction to: a systematic review of the clinical utility of cycle threshold values in the context of COVID-19. Infect. Dis. Ther. 2020;9:587. doi:10.1007/s40121-020-00328-z. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

3. Tom M.R., Mina M.J. To interpret the SARS-CoV-2 test, consider the cycle threshold value. Clin. Infect. Dis. 2020;71:2252–2254. doi:10.1093/cid/ciaa619. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

4. Dahdouh E., Lazaro-Perona F., Romero-Gomez M.P., Mingorance J., Garcia-Rodriguez J. Ct values from SARS-CoV-2 diagnostic PCR assays should not be used as direct estimates of viral load. J. Infect. 2021;82:414–451. doi:10.1016/j.jinf.2020.10.017. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

5. Evans D., Cowen S., Kammel M., O'Sullivan D.M., Stewart G., Grunert H.P., et al. The dangers of using Cq to quantify nucleic acid in biological samples; a lesson from COVID-19. Clin. Chem. 2021 doi:10.1093/clinchem/hvab219. [PubMed] [CrossRef] [Google Scholar]

6. Austrian Federal Ministry for Social Affairs Healthcare, care and consumer protection. Empfehlung für die Gesundheitsbehörden zur Entlassung von COVID-19-Fällen aus der Absonderung. 2021 https://www.sozialministerium.at/dam/jcr:531f7e21-0f53-4180-b214-848e19668b52/Empfehlung_fuer_die_Gesundheitsbeho*rden_zur_Entlassung_von_COVID-19-Faellen_aus_der_Absonderung.pdf [Updated 21 Oct. [Google Scholar]

7. Rhoads D., Peaper D.R., She R.C., Nolte F.S., Wojewoda C.M., Anderson N.W., et al. College of American pathologists (CAP) microbiology committee perspective: caution must Be used in interpreting the cycle threshold (Ct) value. Clin. Infect. Dis. 2021;72:e685–e686. doi:10.1093/cid/ciaa1199. [PubMed] [CrossRef] [Google Scholar]

8. Zeichhardt H., Kammel M., INSTAND Quantitative Bezugsproben zur Verbesserung der Vergleichbarkeit und Bewertung von Laborergebnissen zum Virusgenom-Nachweis von SARS-CoV-2. Informationen zur Testdurchführung und Anwendung der quantitativen Bezugsproben inkl., Anhang Auswertung der Ergebnisse vom 1. Versand (03.11. 2020) und 2. Versand (17.11 2020). https://www.instand-ev.de/fileadmin/uploads/user_upload/Dokumente/Virologie/20210118g_Begleitheft_-_quantitative_Bezugsproben_1_und_2_-_SARS-CoV-2.pdf. (Accessed 15 January 2021) 1-20.

9. Vierbaum L., Wojtalewicz N., Grunert H.P., Lindig V., Duehring U., Drosten C., et al. RNA reference materials with defined viral RNA loads of SARS-CoV-2-A useful tool towards a better PCR assay harmonization. PLoS One. 2022;17 doi:10.1371/journal.pone.0262656. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

10. Abbott Molecular Inc . 53– 608278/R1. Jun 2021. (Package Insert Alinity M SARS-CoV-2 AMP Kit). [Google Scholar]

11. Abbott Molecular Inc . Jun 2020. Package Insert Abbott RealTime SARS-CoV-2, 51-608442/R2. [Google Scholar]

12. Cepheid . vols. 302–5159. Rev. C; Feb 2021. (Package Insert Xpert® Xpress SARS-CoV-2/Flu/RSV Kit). [Google Scholar]

13. Buchta C., Camp J.V., Jovanovic J., Chiba P., Puchhammer-Stockl E., Mayerhofer M., et al. The versatility of external quality assessment for the surveillance of laboratory and in vitro diagnostic performance: SARS-CoV-2 viral genome detection in Austria. Clin. Chem. Lab. Med. 2021;59:1735–1744. doi:10.1515/cclm-2021-0604. [PubMed] [CrossRef] [Google Scholar]

14. Ehret R., Breuer S., Dhein J., Reinhardt B., Obermeier M. Clinical evaluation of the automated Abbott RealTime SARS-CoV-2, alinity m SARS-CoV-2, and alinity m resp-4-plex assays. J. Virol Methods. 2021 doi:10.1016/j.jviromet.2021.114338. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

15. Perchetti G.A., Pepper G., Shrestha L., LaTurner K., Yae Kim D., Huang M.L., et al. Performance characteristics of the Abbott Alinity m SARS-CoV-2 assay. J. Clin. Virol. 2021;140 doi:10.1016/j.jcv.2021.104869. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

16. Hirschhorn J.W., Kegl A., Dickerson T., Glen W.B., Jr., Xu G., Alden J., et al. Verification and validation of SARS-CoV-2 assay performance on the Abbott m2000 and alinity m systems. J. Clin. Microbiol. 2021:59. doi:10.1128/JCM.03119-20. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

17. Rodgers M.A., Olivo A., Harris B.J., Lark C., Luo X., Berg M.G., et al. Detection of SARS-CoV-2 variants by Abbott molecular, antigen, and serological tests. J. Clin. Virol. 2022;147 doi:10.1016/j.jcv.2022.105080. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

18. Lalkhen A.G., McCluskey A. Clinical tests: sensitivity and specificity. Cont. Educ. Anaesth. Crit. Care Pain. 2008;8:221–223. doi:10.1093/bjaceaccp/mkn041. [CrossRef] [Google Scholar]

19. Dierks S., Thiele K., Bohne W., Lugert R., Weig M., Gross U., et al. Comparison and harmonization of different semi-automated and automated qRT-PCR assays in the assessment of SARS-CoV-2. Viruses. 2022;14 doi:10.3390/v14102239. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

20. Kogoj R., Kmetic P., Ostrbenk Valencak A., Fujs Komlos K., Seme K., Sagadin M., et al. Real-life head-to-head comparison of performance of two high-throughput automated assays for the detection of SARS-CoV-2 RNA in nasopharyngeal swabs: the alinity m and cobas 6800 SARS-CoV-2 assays. J. Mol. Diagn. 2021;23:920–928. doi:10.1016/j.jmoldx.2021.05.003. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

21. Kohmer N., Rabenau H.F., Hoehl S., Kortenbusch M., Ciesek S., Berger A. Comparative analysis of point-of-care, high-throughput and laboratory-developed SARS-CoV-2 nucleic acid amplification tests (NATs) J. Virol Methods. 2021;291 doi:10.1016/j.jviromet.2021.114102. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

22. Zhen W., Manji R., Smith E., Wuitschick J., Lucic D., Berry G.J. Evaluation of the alinity m resp-4-plex assay for the detection of severe acute respiratory syndrome coronavirus 2, influenza A virus, influenza B virus, and respiratory syncytial virus. Microbiol. Spectr. 2022;10 doi:10.1128/spectrum.01090-21. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

23. Rondaan C., Gard L., Niesters H.G.M., van Leer-Buter C., Zhou X. COVID or no COVID: interpreting inconclusive SARS-CoV-2 qPCR results in different populations and platforms. J. Clin. Virol. 2023;3 doi:10.1016/j.jcvp.2023.100145. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

24. World Health Organization Clinical management of COVID-19: living guideline, 13 january 2023. https://www.who.int/publications/i/item/WHO-2019-nCoV-clinical-2023.1 [Updated. [PubMed]

25. Public Health England. Understanding cycle threshold (Ct) in SARS-CoV-2 RT-PCR. PHE gateway number: GW-1651. [Updated] https://assets.publishing.service.gov.uk/government/uploads/system/uploads/attachment_data/file/926410/Understanding_Cycle_Threshold__Ct__in_SARS-CoV-2_RT-PCR_.pdf Accessed May 19, 2023 1-12.

Articles from Heliyon are provided here courtesy of Elsevier

Clinical comparison of three SARS-CoV-2 nucleic acid amplification tests for routine diagnostic testing (2024)
Top Articles
Facebook Marketplace Marrero La
Snow Bunny Meaning: What Does This Slang Term Really Refer To? - English Study Online
Barstool Sports Gif
Busted Newspaper Zapata Tx
Jazmen Jafar Linkedin
Katmoie
Stl Craiglist
ds. J.C. van Trigt - Lukas 23:42-43 - Preekaantekeningen
Osrs Blessed Axe
Raid Guides - Hardstuck
Voyeuragency
Craiglist Galveston
Conan Exiles Thrall Master Build: Best Attributes, Armor, Skills, More
Ou Class Nav
Jellyfin Ps5
Puss In Boots: The Last Wish Showtimes Near Cinépolis Vista
Atlases, Cartography, Asia (Collection Dr. Dupuis), Arch…
Greyson Alexander Thorn
Manuela Qm Only
R/Airforcerecruits
Garden Grove Classlink
Jersey Shore Subreddit
LG UN90 65" 4K Smart UHD TV - 65UN9000AUJ | LG CA
Ellafeet.official
How to Use Craigslist (with Pictures) - wikiHow
Craigslist Free Puppy
Teenbeautyfitness
Haley Gifts :: Stardew Valley
Tamil Play.com
Ark Unlock All Skins Command
Jr Miss Naturist Pageant
Rise Meadville Reviews
Log in or sign up to view
Metro 72 Hour Extension 2022
Ippa 番号
Whitehall Preparatory And Fitness Academy Calendar
Smith And Wesson Nra Instructor Discount
Mvnt Merchant Services
Cal Poly 2027 College Confidential
Craigslist Pa Altoona
Arnesons Webcam
Love Words Starting with P (With Definition)
Babykeilani
Studentvue Calexico
Zeeks Pizza Calories
Crigslist Tucson
Craigslist Pets Charleston Wv
The Hardest Quests in Old School RuneScape (Ranked) – FandomSpot
Zom 100 Mbti
Dcuo Wiki
Southern Blotting: Principle, Steps, Applications | Microbe Online
Obituary Roger Schaefer Update 2020
Latest Posts
Article information

Author: Pres. Carey Rath

Last Updated:

Views: 6307

Rating: 4 / 5 (61 voted)

Reviews: 92% of readers found this page helpful

Author information

Name: Pres. Carey Rath

Birthday: 1997-03-06

Address: 14955 Ledner Trail, East Rodrickfort, NE 85127-8369

Phone: +18682428114917

Job: National Technology Representative

Hobby: Sand art, Drama, Web surfing, Cycling, Brazilian jiu-jitsu, Leather crafting, Creative writing

Introduction: My name is Pres. Carey Rath, I am a faithful, funny, vast, joyous, lively, brave, glamorous person who loves writing and wants to share my knowledge and understanding with you.